Multi-Element Silicon Drift Detectors

For a variety of beamline applications, Mirion offers a standard array configuration with 7 elements of 80 mm² active area, as well as customizable systems.

7-Element SDD Array

Description:

The X-PIPS Detector is a spectroscopy sub-system sensitive to X-rays and low-energy gamma rays. It comprises a hermetically sealed silicon drift detector (SDD) array with a low noise CMOS reset type preamplifier and Peltier cooler, a HV bias supply and a temperature controller. The detector elements and CMOS preamplifiers are cooled and temperature regulated, ensuring stable operation in changing environmental conditions. The Beryllium entrance window is 1 mil (25 µm) thick.

The preamplifiers have a fast reset mechanism providing short dead time and excellent count rate performance. The reset can be synchronous or asynchronous (user selectable).

The high performance SDD combined with a CMOS preamplifier provides very fast and low noise response, which results in extremely good energy resolution with fast peaking times. The signal rise time is well below 50 ns which makes for excellent energy resolution at high count rates.

The energy resolution is guaranteed within an ambient temperature range of +10 °C to + 30 °C with the default factory settings.

The X-PIPS Detector is equipped with a multilayer collimator providing excellent peak to background ratios and few spurious peaks.

The Peltier cooled detector assembly has a cool down time of just a few minutes.

Key Features

Detector System Includes:

  • Seven Silicon Drift Detectors (SDDs)
  • 1 mil (25 µm) Be Window
  • CMOS Preamplifiers
  • HV Bias Supply
  • Temperature controlled Peltier Cooler

Performance

  • Collimated Active Area – 7x80 mm2
  • Thickness – 0.5 mm
  • Resolution <145 eV (FWHM)*
  • Energy Range – 1 to 30 keV

Applications

  • X-ray Absorbtion Spectroscopy (EXAFS, XANES)
  • X-ray Fluorescence
  • X-ray Diffraction
  • Mössbauer Spectroscopy
  • Densitometry

Custom Configurations

Key Features

  • Peltier or Cryo-cooled
  • Up to 7 (with Peltier) or 13 (with Cryo-cooler) elements
  • Planar or focused configurations
  • CMOS based preamplifiers
  • Integrated HV bias supply

Performance

  • Collimated Active Area –
    30-80 mm2 per detector
  • Thickness – 0.5 mm
  • Optimum resolution <135 eV (FWHM)
  • Energy Range – 1 to 30 keV
  • P/B ratio >10,000:1

Applications

  • X-ray Absorbtion Spectroscopy (EXAFS, XANES)
  • X-ray Fluorescence
  • X-ray Diffraction
  • Mössbauer Spectroscopy
  • Densitometry
EXAMPLES

Example 1:

  • 7x50 mm2 collimated elements
  • Cryo-cooled
  • 1.5 W Pulse-Tube cooler
    • High reliability and long life (>10 years)
    • Low power consumption (<50 W)
    • Low vibration levels
  • Air or water cooled heat sink
 

Example 2:

  • 7x50 mm2 collimated elements
  • Focused arrangement
  • Peltier cooled
  • Air or water cooled heat sink
 

Example 3:

  • 8x80 mm2 collimated elements
  • Focused arrangement
  • Cryo-cooled
 
 

PIPS® is a registered trademark of Canberra Industries, Inc.